Home > News > Product News > General Safety Regulations for Summer Laboratories

General Safety Regulations for Summer Laboratories

Jul. 22. 2025

General Safety Regulations for Summer Laboratories

Special measures for high temperature warning days

General Safety Regulations for Summer Laboratories

When the ambient temperature exceeds 35℃:

Shorten the continuous operation time of the equipment (single operation ≤4h)

Prohibit solvents with boiling points below 60℃ (such as ether and acetone)

Equipped with dry ice emergency kit (to handle sudden refrigeration failures)


Preventive maintenance of pre equipment components

Refrigerant pressure inspection, high pressure ≤1.8MPa, low pressure ≥0.3MPa

Circulating pipeline inspection, no crystallization, discoloration or abnormal expansion

Electrical wiring inspection, insulation resistance>5MΩ


Emergency response prompt:

When there is a leakage of heat transfer oil or refrigerant splashing,

1. Immediately activate the ventilation system

2. Cut off the main power supply of the equipment

3. Cover the leakage point with inert absorbent cotton (no water flushing!)

4. Contact professional hazardous chemical disposal institutions


This guide is applicable to scenarios such as pharmaceuticals, chemical engineering, and university research and development. Regular implementation can reduce the risk of equipment failure by 80% during high temperature seasons. It is recommended that laboratory administrators print and post it in the equipment operation area.